Polyolefins are macromolecules formed by the polymerization of olefin (alkene) monomer units. The most common polyolefins are polypropylene (PP) and polyethylene (PE) homopolymers. These polymers have a wide range of applications -- household goods, industrial products, and specialty medical products. Their properties can be further tuned to perform beyond the typical applications. For example, Ethylene Propylene Diene Monomer (EPDM) is a popular olefin elastomer used in automotive sealant and electrical insulation. The purpose of this App Note is to present the range of Intrinsic Viscosities (IV) achievable with an HMJ Y501 Dual Capillary High Temperature Relative Viscometer (HTDSV) system. The analysis conditions are listed below.
Solvent |
Decalin |
Sample Concentration |
0.0025 g/dL to 0.5 g/dL |
Prep Temperature |
135C - 150C |
Dissolution Time |
60 minutes - Overnight |
Analysis Temperature |
30C |
The summary table below clearly shows excellent correlation between HT-DSV measured IV data as compared to calculated or expected IV values.
PE Samples |
Measured IV (dL/g) |
Calculated/Expected IV |
NIST-1475a |
1.1645 @ 140C |
1.180 @ 130C (NIST COA) |
HMW-PE |
13.65 @ 150C |
Unknown (Calculated MW = 1,600,000) |
UHM-PE (3,100,000) |
13.65 @ 150C |
Calculated IV = 23 |
UHM-PE (3,400,000) |
23 @ 135C |
Calculated IV = 24 |
UHM-PE (4,000,000) |
26 @ 135C |
Calculated IV = 26 |
UHM-PE (8,000,000) |
In-Progress |
Calculated IV = 43 |
In addition to the great IV range demonstrated above, the relative viscometer presents a great opportunity to automate this particularly laborious task. Our fully automated High Temperature Vortex DSV system is ideal for this difficult and hazardous measurement.