Spraylink Spray
|
Spraylink Spray Particle Size Analyzer
The Spraylink system delivers routine, accurate and precise spray particle size measurement in the size range of 0.1 µm to 2080 µm with no need for a lens change. It integrates an optimized high-sensitivity detection and holographic signal processing system, combined with high-concentration compensation technology, which greatly reduces the effect of multiple diffraction during the measurement of high-concentration spray of up to 90% obscuration.
The Spraylink Optical System IP65-rated Transmitter and Receiver Units are mounted on an optical bench. It includes a multi-element silicon photo-diode array, rapid data acquisition system, optical auto-align stage, one output TTL trigger to control external devices, one input TTL trigger to synchronize measurements with other devices and USB/optical fiber connection to a computer. Data Acquisition Rate of up to 10kHz.
Spraylink software provides a user-friendly graphical interface for controlling all aspects of a spray measurement with full measurement automation. Results include high concentration analysis, a real-time display providing a size-history of the measured spray event, data averaging, data searching and sorting facilities. The software can also be operated as a stand-alone utility for post-measurement analysis work away from the instrument.
The Spraylink system delivers routine, accurate and precise spray particle size measurement in the size range of 0.1 µm to 2080 µm with no need for a lens change. It integrates an optimized high-sensitivity detection and holographic signal processing system, combined with high-concentration compensation technology, which greatly reduces the effect of multiple diffraction during the measurement of high-concentration spray of up to 90% obscuration.
The Spraylink Optical System IP65-rated Transmitter and Receiver Units are mounted on an optical bench. It includes a multi-element silicon photo-diode array, rapid data acquisition system, optical auto-align stage, one output TTL trigger to control external devices, one input TTL trigger to synchronize measurements with other devices and USB/optical fiber connection to a computer. Data Acquisition Rate of up to 10kHz.
Spraylink software provides a user-friendly graphical interface for controlling all aspects of a spray measurement with full measurement automation. Results include high concentration analysis, a real-time display providing a size-history of the measured spray event, data averaging, data searching and sorting facilities. The software can also be operated as a stand-alone utility for post-measurement analysis work away from the instrument.
> 10kHz maximum acquisition rate, perform 10000 measurements per second and capture the dynamics of size change in every 0.1ms interval
> Unique lens configuration and optimized optics enable measurements over large working ranges
> Measurement across a wide range (0.1-2080um) with no need for optics change
> The newest algorithm ensures accurate particle size distributions can be measured at high obscuration
> Rapid auto-alignment
> Lens protection with air purging
> Customizable optical bench arrangement
> Measurement synchronization via flexible triggering options
> Stable construction and full metal cover and optical bench design
>I nstantaneous playback of size history and frame-by-frame event analysis
> Analysis of the dynamic changes of particle size distribution with time
> Overlay of results at different time point
> Unique lens configuration and optimized optics enable measurements over large working ranges
> Measurement across a wide range (0.1-2080um) with no need for optics change
> The newest algorithm ensures accurate particle size distributions can be measured at high obscuration
> Rapid auto-alignment
> Lens protection with air purging
> Customizable optical bench arrangement
> Measurement synchronization via flexible triggering options
> Stable construction and full metal cover and optical bench design
>I nstantaneous playback of size history and frame-by-frame event analysis
> Analysis of the dynamic changes of particle size distribution with time
> Overlay of results at different time point
Specification |
|
Measuring Principle |
Laser light scattering |
Optical models |
Mie theory and Fraunhofer scattering,including high concentration technique |
Size range |
0.1 μm - 2080 μm, No need for lens change and calibration |
Detection System |
Detectors array with area compensation and gridded holder |
Scope |
0.3 μm : 150 mm; 7 μm : 1800 mm |
Light Source |
Max. 20 mW, 638 nm Solid-state laser with thermostat integrated |
Density |
Maximum shading degree: 90% (depends on size range) |
Optical alignment system |
Smart rapid align system with full automation |
External synchronization |
TTL trigger output |
Enclose rating |
IP 65 for transmitter and receiver modules |
Accuracy |
Better than 0.6% (The mean size of NIST latex standard) |
Repeatability |
Better than 0.5% (Sample and sample preparation dependent) |
Optical system weight |
28 kg |
Optical system dimensions |
1100 mm x 460 mm x 538 mm |